attoPUBLICATIONS attoDRY | Page 15

RT version

RT-SWD( RT-USWD) vacuum shroud for use with room temperature objective and positioners

PRODUCT KEY FEATURES
• standard configuration for use with external RT objective and xyz positioners(+ optional scanner)
• Ultra-short working distance upgrade available( 1 mm / 2 mm)
• compatible with 30 mm & 60 mm cage system
• optional sample holder with 12 optional electrical contacts; temp. sensor & heater included
• feedthrough ring with blind flanges included, DC / SMA / fiber feedthroughs optional
3 4
9
„ ZOOM out“
1
10
1
6
2
5
1
2 3 4
5 attoDRY800 RT-SWD vacuum shroud for use with room temperature objective; default min. working distance < 4 mm( others on request, see also USWD option with 2 mm( 1 mm))
cold shield with top lid adjustable for min. working distance vacuum top window, 25 mm x 1.5 mm, fused silica( others on request)
cold shield top window, 12.7 mm x 0.375 mm, fused silica( others on request); default cold shield clear aperture 9 mm( 3, 6 11, 14 mm on request)
cold plate
6
7 8
9 10
ANPxy311 / ANPz102 positioners + ATC100 / Si thermal link with integrated Si diode & heater + optional ANSxy100
ASH / SWD sample holder for attoDRY800
ASH / PCB / 12mini for ASH / SWD with 12 electrical contacts wired to twisted pairs
30 mm x 30 mm & 60 mm x 60 mm cage system mounting threads centered around top window; 4x cage rod mounting adapters for quick removal included
feedthrough ring with blind flanges included, DC / SMA / fiber feedthroughs optional
attoDRY
Closed-cycle Cryostats