attoPUBLICATIONS attoDRY | Page 14

Basic( Standard Shroud) basic vacuum shroud for use with a room temperature objective

RT version
PRODUCT KEY FEATURES
• included with attoDRY800 base configuration at no additional cost!*
• corresponds to the classical flow cryostat configuration
• 88xM6 breadboard mounting holes
• compatible with 30 mm & 60 mm cage system
• optional sample holder with manually adjustable height, and 12 optional electrical contacts
• feedthrough ring with blind flanges included, DC / SMA / fiber feedthroughs optional
• not compatible with positioners, see e. g. RT-SWD or others
* no additional cost only in case no other vacuum shroud is purchased
8 9
3
7
10
1
2
4
5
6
„ ZOOM out“
1
2 3 4
5 attoDRY800 basic vacuum shroud for use with room temperature objective( max. objective outer diameter 36 mm)
cold shield with top lid adjustable for min. working distance vacuum top window, 25 mm x 1.5 mm, fused silica( others on request)
default min. working distance of 3.5 mm( others on request) cold shield top window, 12.7 mm x 0.5 mm, fused silica( others on request); default cold shield clear aperture 9 mm( 3, 6 11, 14 mm on request)
cold plate
6 7
8 9
10
ASH / Zvar sample holder with +/-7 mm manual z range
ASH / PCB / 12mini for ASH / Zvar with 12 electrical contacts wired to twisted pairs
88x M6 breadboard mounting holes on lid( 12.5 mm grid)
30 mm x 30 mm & 60 mm x 60 mm cage system mounting threads centered around top window; 4x cage rod mounting adapters for quick removal included
feedthrough ring with blind flanges included, DC / SMA / fiber feedthroughs optional