PRODUCT KEY FEATURES
• standard configuration for use with internal high vacuum
objective and xyz positioners (+ optional scanner)
• compatible with 30 mm & 60 mm cage system
• optional sample holder with 12 optional electrical contacts;
temp. sensor & heater included
• feedthrough ring with blind flanges included, DC/SMA/fiber
feedthroughs optional
10
3
1
9
3
2
4
8
11
7
2
6
5
„ZOOM out“
1 attoDRY800 HV objective vacuum shroud for use with high
vacuum compatible objective
2
3 cold shield with top lid adjustable for min. working distance
4
5
6 4x side vacuum windows 40 mm x 4 mm
top vacuum window 25 mm x 1.5 mm, cold shield top window
12.7 mm x 0.375; window material fused silica (others on request)
cold plate
ANPxy311/ANPz102 positioners + ATC100/Si thermal link
with integrated Si diode & heater + optional ANSxy100 scanner
7
8
9
ASH/SWD sample holder for attoDRY800
ASH/PCB/12mini for ASH/SWD with 12 electrical contacts wired to twisted pairs
HV compatible, RMS standard objective with 45 mm length; several objectives
available on request (NA = 0.75 .. 0.95)
10 30 mm x 30 mm & 60 mm x 60 mm cage system mounting threads centered around
top window; 4x cage rod mounting adapters for quick removal included
11 feedthrough ring with blind flanges included, DC/SMA/fiber
feedthroughs optional