Performance Data * | |
temperature range |
4 .. 300 K |
cool down time of sample |
approx . 2 h ( depending on insert ) |
cool down time of system ( system without magnet ) |
approx . 5 .. 10 h ( unattended ) |
cool down time of system ( system incl . 9 T magnet ) |
approx . 10 .. 15 h ( unattended ) |
temperature stability |
< ± 10 mK expected ( 4 .. 50 K ) |
< ± 25 mK guaranteed ( 4 .. 50 K ) | |
cooling power at sample location |
> 5 mW @ 5 K |
Closed-cycle cooler | |
nominal cooling power ( 4.2 K ) |
> 900 mW |
power consumption |
max . 9.0 kW , 7.2 kW steady state |
cooling of compressor |
water cooling ( requires local infrastructure ) |
Dimensions | |
cryostat ( width x depth x height ) |
1120 x 640 x 1050 mm ³ ( depending on magnet choice ) |
optional electronics rack ( width x depth x height ) |
640 x 640 x 1050 mm ³ |
Options | |
superconducting magnet |
9 T |
bipolar magnet power supply |
included ( with optional magnet ) |
temperature controller |
included |
pumping kit |
included |