ANC300 Plug-In Modules open loop control for attocube positioners

PAGE 395
ANC300
The ANC300 offers a completely modular design with up to seven slots for dedicated stepping( ANM150), scanning( ANM200) or combined stepping and scanning( ANM300) modules.
ANM150
stepping module
Output Signals
stepping- voltage range
0.. 150 V
stepping- frequency range
0.. 10 kHz
stepping- maximum current
100 mA, 4.5 A peak( max. 1 ms)
resolution of signal generation
2.3 mV( 16 bit)
output noise
< 5 mVpp( 20 MHz bandwidth)
maximum capacitive load
5 μF
zero voltage output in GND mode
connected to chassis ground directly
output connectors
BNC, 50 Ohm
ANM200 scanning module
Output Signals
scanning- output voltage range
0.. 150 V
scanning- gain
15
scanning- maximum current
100 mA
scanning- offset
programmable offset
scanning- power bandwidth
DC.. 500 Hz( 1 µ F load)
scanning- small signal bandwidth
DC.. 100 kHz( 1 µ F load) via DC input, up
to 500 kHz via AC input
scanning- slew rate
800 V / ms( 1 µ F load)
resolution of signal generation
2.3 mV( 16 bit)
output noise
< 5 mVpp( 20 MHz bandwidth)
output filter
additional switchable output filter
( 1.6, 16, 160, 1600 Hz)
maximum capacitive load
5 µ F
zero voltage output in GND mode
connected to chassis ground directly
output connectors
BNC, 50 Ohm
Input Signals
DC in voltage range
-10..+ 10 V
AC in voltage range
max. ± 5 V AC
coupling
DC, programmable offset
input resistance
10 kOhm
input connectors
BNC, 50 Ohm
ANM300 stepping & scanning module
Output Signals
stepping- voltage range
0.. 150 V
stepping- frequency range
0.. 10 kHz
stepping- maximum current
100 mA, 4.5 A peak( max. 1 ms)
scanning- output voltage range
0.. 150 V
scanning- gain
15
scanning- maximum current
100 mA
scanning- offset
programmable offset
scanning- power bandwidth
DC.. 500 Hz( 1 µ F load)
scanning- small signal bandwidth
DC.. 100 kHz( 1 µ F load) via DC input, up
to 500 kHz via AC input
scanning- slew rate
800 V / ms( 1 µ F load)
resolution of signal generation
2.3 mV( 16 bit)
output noise
< 5 mVpp( 20 MHz bandwidth)
maximum capacitive load
5 μF
zero voltage output in GND mode
connected to chassis ground directly
output connectors
BNC, 50 Ohm
Input Signals
DC in voltage range
-10..+ 10 V
AC in voltage range
max ± 5 V AC
coupling
DC, programmable offset
input resistance
10 kOhm
input connectors
BNC, 50 Ohm