Specifications attoRAMAN

General Specifications type of instrument
sensor head specifics
Confocal Unit configuration
quick-exchange of optical components
LT- compatible objective free-beam based room temperature optics head coupled to low temperature objective and ultra-high transmission spectrometer
unique low temperature compatible achromatic objectives with high numerical aperture , optimized for different wavelength ranges
compact and modular design , two or more optical channels standard configuration : 1 excitation channel , 1 detection channel
beamsplitters , filter mounts for up to 4 filters / polarizers ( 1 " diameter ), optional piezoelectric rotator with filter mount
LT-APO / VIS , LT-APO / VISIR , LT-APO / NIR ( see accessory section for more information )
inspection unit
sample imaging with large field of view : ~ 75 μm ( attoDRY ), ~ 56 μm ( attoLIQUID )
Illumination
excitation wavelength range
400 .. 1000 nm
default : 532 nm ( others on request )
light source
dedicated Raman laser , single mode fiber coupled
light power on the sample
typically 1 pW .. 10mW
optical filter
laser line filter
Detection
detection mode
2D Raman images , time and single point Raman spectra
spectrometer
ultra-high transmission spectrometer , f = 300 mm
total optical transmission
greater 60 % at 532 nm
filters
dichroic mirror & edge filter for signal detection as close as 90 cm-1 to the laser line
gratings
typ . 600 / mm and 1800 / mm grating
spectral resolution
1 cm-1 at 1800 / mm grating
CCD camera
back-illuminated CCD , peltier-cooled to -60 ° C at 20 ° C room temperature , 1024x127 pixels , 90 % quantum
efficiency at 532 nm , 100 kHz readout converter
Sample Positioning
total travel range
5 x 5 x 5 mm ³ ( open loop )
step size
0.05 .. 3 μm @ 300 K , 10 .. 500 nm @ 4 K
fine scan range
50 x 50 μm ² @ 300 K , 30 x 30 μm ² @ 4 k ( open loop )
sample holder
ASH / QE / 0 quick exchange sample holder and integrated heater with calibrated temperature sensor
Suitable Operating Conditions
temperature range
1.5 K .. 300 K ( dependent on cryostat ); mK compatible setup available on request
magnetic field range
0 .. 15 T + ( dependent on magnet )
operating pressure
designed for He exchange gas ( vacuum compatible version down to 1E-6 mbar on request )
Suitable Cooling Systems
titanium housing diameter
48 mm
bore size requirement
designed for a 2 " ( 50.8 mm ) cryostat / magnet bore
compatible cryostats
attoDRY1000 / 1100 / 2100
attoLIQUID1000 / 2000 ( attoLIQUID3000 / 5000 on request )
Complete microscope stick attoRAMAN